Author: Lee Publish Time: 2026-11-20 Origin: Site
The pin leg spacing measured in the free state of the nickel-titanium alloy bone fixation pin may differ from the reading when it is mounted within the implant device and in a constrained state. To ensure reproducibility of the measurement results, the drawings must clearly indicate the center distance of the pin legs or the inner clearance, the cross-section at which the measurement is taken, and the condition of the fixation pin at the time of measurement. Clear edge acceptance criteria are also required for the transition area from the pin tip to the inside of the pin leg.
Quick response
CNC machining can be used to manufacture or finish the local features of nickel-titanium alloy bone fixation pins. This includes the contour of the pin legs, the rounding of transitions, and the contact surfaces of the implant device. Depending on the shape of the parts and the form of the raw materials, the complete process may include molding, wire electrical discharge machining, shape setting, and surface finishing. During the initial inspection, the pin leg spacing must be measured in the cross-section and condition of the parts specified in the drawings, and critical dimensions and edge conditions must be re-inspected after final finishing.
Through CNC machining, the research and development team for orthopedic instruments can verify various design proposals before introducing dedicated mass production jigs. Engineers can modify the contour of the pin legs, the rounding of the bridge section, or the contact surfaces of the implant device and confirm fit with new sample parts. This is particularly useful for projects where bone fixation pins and implant devices are developed simultaneously.
When transitioning to trial production, the procurement department must confirm that suppliers can consistently deliver based on the same version of the drawings, material conditions, and inspection requirements. Nickel-titanium alloy is sensitive to processing history. Even if the CAD model is the same, changes in the state of the raw materials, shape setting, or surface finishing processes can affect the product. The quotation must specify material specifications, the condition of the parts at final inspection, and the responsible parties for machining, shape setting, and finishing.Understanding the manufacturing capabilities of Dawon Precision's medical device parts
If the drawings require local planes, specified rounding, or positioning surfaces on the pin legs, the adoption of CNC milling can be evaluated. During the development period, these changes can be reflected in the next sample parts by modifying the machining programs and clamping schemes.
The overall process depends on the shape of the blank and the requirements of the finished product.The wire rod scheme can first obtain the basic shape through forming, and then process local features as needed.The sheet metal scheme considers obtaining the main contour using wire cutting, and then completes the finishing process as specified in the drawings. When formulating the process, it is necessary to clarify the dimensions to be inspected in the final state and to establish quality requirements for the wire cutting surface, nail tips, and inner rounded corners.Check out the CNC machining services of Dawon Precision.
The specifications of the UNITUS™ Staple System released by Zimmer Biomet range from 8 × 8 mm to 30 × 20 mm. This set of specifications indicates the width of the bridge part × the length of the staple legs. The actual spacing of the staple legs and their tolerances are specified in the drawings of each part.
Parts or features | Key points for drawing review |
Two-leg bone fixation staple | Width and contour of the bridge part, length and cross-section of the staple legs, and spacing of the staple legs at the specified cross-section |
Staple legs with machining characteristics | Position and shape of flat, groove, or stop-retreat characteristics, as well as edge quality requirements |
Transition area from the bridge part to the staple legs | Inner curvature, contour tolerances, and the accessibility of finishing and inspection |
Clamp or positioning insert for the implant | Contact surfaces, positioning depth, and the position of the fixing pins in a constrained state |
Inspection jig | Positioning surfaces, contact points, and the specified constraints applied to the pin legs |
When determining the condition of parts at the time of acceptance for jigs or implants, it is necessary to provide a version of the drawings corresponding to the estimate documents, allowing the supplier to re-measure under the same conditions using the specified matching parts.
If the pin legs converge inward or spread outward, the spacing between the part near the bridge and the part near the pin tips will differ. The drawings must specify the distance from the reference to the measurement section and clarify whether the measurement is the center distance of the pin legs or the pure distance between the inner surfaces of the two pin legs. If both conditions need to be inspected, the results in the free state and constrained state must be recorded separately. For elongated pin legs, non-contact optical measurement can be used to avoid displacement of the pin legs due to contact from the measurement head, assuming that the measurement capability meets the tolerance requirements.
The roughness measurements reflect the texture of the measured surface. Burrs at the tip of nails, edge dulling, and internal rounding should be inspected according to their respective requirements. The drawings should explicitly indicate important edges and specify the acceptable condition of edges, or agree on limit samples for comparison during inspection. If the finish may alter the pure distance of adjacent contours or the inside of nail legs, these features should be included in the final inspection.
When processing narrow nail legs or small inner angles, the condition of the tool, the support of the parts, and the cutting participation amount of the tool can affect the quality of the edge. During trial cutting, the contour dimensions and edge conditions can be synchronized and recorded at the same position. Even if the dimensions are within tolerance, if burrs begin to occur at the nail tip or the tool exit position, the tool and part support should first be checked, and then a decision should be made on whether to adjust the cutting parameters.
The public 510(k) summary of the restor3d Extremity Staple System (K231458) introduces bone fixation nails made of nickel-titanium alloy with various bridge lengths and nail leg lengths. The submitted device adopts a circular cross-section nail leg, while the main control device uses a square cross-section nail leg. The summary lists static and dynamic bending tests, pull-out tests, and cyclic corrosion tests.
At the time of procurement, the cross-section of the nail leg, the geometric fit of the implant device, and the nominal spacing should be described together in the requirements. Even if the nominal spacing is the same, two designs may require different measurement criteria and edge inspection methods. This public document introduces the structure of the device and the verification tests. The following cutting data is from another published NiTi processing study.
In "Machining Surface Roughness and Cutting Force during Milling of NiTi Alloy," the researchers used a coated carbide end mill with a diameter of 6mm, under conditions of a spiral angle of 30°, a cutting speed of 45m/min, and a feed per tooth of 0.02mm, measuring a surface roughness Sa of 0.301µm.
This dataset demonstrates how to evaluate surface quality by combining tool geometry and cutting parameters. During the trial cutting of bone fixation nails, parameters need to be determined based on the actual blade diameter, the material condition of the nickel-titanium alloy, the rigidity of the nail leg, and the clamping method. The Sa reported in the paper is the arithmetic mean height of the measurement area. If the arithmetic mean roughness Ra of the contour is specified in the drawing, both should be measured individually using an agreed-upon method.
The table below is an example of the process, and all numerical values and readings are hypothetical, illustrating how to record measurement conditions, readings before and after finishing, and final acceptance results in the initial report. Actual dimensions, tolerances, and inspection conditions are based on the OEM drawings and inspection plans.
Inspection Items | Examples of drawing requirements or inspection conditions | Examples of initial results |
Clearance on the inside of the nail foot in a free state | Measured at a section 3.00 mm from the specified reference surface of the bridge, with a requirement of 12.00 ±0.05 mm | 11.98 mm before finishing, 12.01 mm after finishing |
Inner transition radius from the bridge to the nail foot | Check the area specified in the drawing, with a requirement of R0.50 ±0.05 mm | After finishing R0.48 mm |
Condition of the edge at the tip of the nail | Magnify 20 times under specified lighting conditions and compare with the agreed limit sample | No burrs exceeding the acceptance criteria of the limit sample were found in the designated area |
Compatibility of the implant instruments | Use the specified version of the implant instruments to confirm that the bridge part is in the designated position under constrained conditions | The initial sample is completely in the designated position |
In this sample, the clearance inside the nail leg increased from 11.98 mm to 12.01 mm, with a change of 0.03 mm. The final reading remains within the sample acceptance range of 11.95–12.05 mm. Measurements must be taken under the same cross-section and free state conditions before and after finishing, allowing for comparison of reading values. The initial sample report must also include the drawing version, measurement method, material lot number, mutually agreed shape settings, and finishing records.
The initial sample report describes whether the features of the inspection target conform to the drawings. The functional testing, acceptance criteria, and verification responsibilities of the finished implant are defined by the OEM verification plan. Procurement should list the items for which the supplier is responsible in the estimate, aiding in the confirmation of inspection man-hours, outsourcing processes, and delivery times.
Dawang Precision has 27 years of manufacturing experience and more than 400 advanced machine tools, including Röders and Mazak five-axis machining centers.Our five-axis CNC machining capabilities can achieve partial critical dimension tolerances of ±0.003 mm and surface roughness of Ra 0.4 µm, with inspections conducted using ZEISS three-dimensional measuring machines. The machining range, target tolerances, and inspection methods for the nickel-titanium alloy bone fixation pin project will be confirmed after reviewing the specific drawings.Understanding Dawang Precision's five-axis CNC machining capabilities
During the estimation phase, we collaboratively review the drawings for bone fixation pins, implant bonding surfaces, and inspection fixtures, identifying features that affect clamping and inspection labor, as well as areas requiring specialized process support. Once the bonding surfaces are determined, we can review the manufacturing plan for the bone fixation pins in parallel with the implant samples and inspection fixtures.
Please send the STEP model and PDF engineering drawings to Dawang Precision's engineering team. Also provide the material specifications and supply status for the nickel-titanium alloy, the measured cross-section and condition of the pin leg spacing, the requirements for critical edges and surface quality, as well as the drawings and expected quantities for the implants.